Six-lead device superior to single-lead smartwatch ECG in atrial fibrillation detection - 02/09/22
Résumé |
This was a head-to-head comparative study on different electrocardiogram (ECG)-based smartwatches and devices for atrial fibrillation detection. We prospectively included 220 patients scheduled for electrical cardioversion and recorded ECGs with 3 different devices (Withings Move ECG, Apple Watch 5, Kardia Mobile 6-leads) as well as the standard 12-lead ECG (gold standard), both before and after cardioversion.
All atrial fibrillation detection algorithms had high accuracy (sensitivity and specificity: 91–99%) but were hampered by uninterpretable recordings (20–24%). In cardiologists’ interpretation, the 6-lead device was superior (sensitivity 99%, specificity 97%) to both single-lead smartwatches (P < .05) for atrial fibrillation detection.
Le texte complet de cet article est disponible en PDF.Graphical abstract |
Six-lead device superior to single-lead smartwatch ECG in atrial fibrillation detection 220 patients (age 70 ± 10 years, female 35%) scheduled for electrical cardioversion.
Plan
Vol 253
P. 53-58 - novembre 2022 Retour au numéroBienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.
Déjà abonné à cette revue ?