Atrial fibrillation detection using ambulatory smartwatch photoplethysmography and validation with simultaneous holter recording - 05/04/22
Abstract |
Background |
Atrial fibrillation (AF)-associated embolic stroke is preventable, and AF detection may help to prevent stroke in subjects with paroxysmal AF. We aimed to evaluate the AF detection performance of smartwatch photoplethysmography (PPG) and the feasibility of ambulatory monitoring for AF detection in the daily life.
Design and methods |
Consecutive subjects who underwent ambulatory Holter electrocardiogram (ECG) monitoring for AF detection or AF burden evaluation were enrolled. The participants underwent 24 hours of simultaneous Holter ECG monitoring and continuous PPG recording using a Garmin smartwatch. The PPG signals were processed for noise rejection, beat detection, beat labeling, and rhythm labeling for each 5-minute segment. The accuracy of the PPG AF detection was calculated using the corresponding simultaneous Holter ECG as the AF diagnostic standard.
Results |
Among the 200 available participants, 112 participants (56%) developed AF (the AF group). The sensitivity, specificity, and positive predicted value of AF detection in participants were 97.3%, 88.6%, and 91.6%, respectively. The area under the receiver operating characteristic curve was 0.90. When the performance was analyzed in these 5-minute segments, the sensitivity, specificity, and positive predicted values of AF detection were 97.1%, 86.8%, and 89.7%, respectively.
Conclusions |
This study demonstrated the feasibility of ambulatory monitoring for AF detection using a commercial smartwatch in daily life. A smartwatch may be an alternative screening tool to standard ambulatory Holter monitoring.
Le texte complet de cet article est disponible en PDF.Graphical abstract |
Plan
Vol 247
P. 55-62 - mai 2022 Retour au numéroBienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.
Déjà abonné à cette revue ?