Are loose-fitting powered air-purifying respirators safe during chest compression? A simulation study - 28/05/21

Doi : 10.1016/j.ajem.2020.03.054 
Soo Hyun Park, MD a, 1, Sung Yeon Hwang, MD a, 1, Guntak Lee, MD a, Jong Eun Park, MD a, Taerim Kim, MD a, Tae Gun Shin, MD a, Min Seob Sim, MD a, Ik Joon Jo, MD a, Seonwoo Kim, PhD b, Hee Yoon, MD a,
a Department of Emergency Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea 
b Statistics and Data Center, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea 

Corresponding author at: Department of Emergency Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, 81 Irwon-ro, Gangnam-gu, Seoul 06351, Republic of Korea.Department of Emergency MedicineSamsung Medical CenterSungkyunkwan University School of Medicine81 Irwon-roGangnam-guSeoul06351Republic of Korea

Abstract

Background

The application of appropriate personal protective equipment for respiratory protection to health care workers is a cornerstone for providing safe healthcare in emergency departments. We investigated the protective effect and usefulness of loose-fitting powered air-purifying respirators (PAPRs) during chest compression.

Methods

This was a single-center simulation study performed from May 2019 to July 2019 in a tertiary hospital. We measured the concentrations of ambient aerosol and particles inside the loose-fitting PAPR during chest compression, and this ratio was set as the simulated workplace protecting factor (SWPF). According to the National Institute for Occupational Safety and Health regulations, the assigned protection factor (APF) of loose-fitting PAPRs is 25. Thus, the loose-fitting PAPRs were assumed to have a protective effect when the SWPF were ≥ 250 (APF × 10). We measured the SWPF of PAPR in real time during chest compression and also investigated the problems encountered during its use.

Results

Ninety-one participants (median age 29 [interquartile range (IQR): 26–32] years; 74% female) completed the simulation. None of the participants failed with SWPF below 250 during three sessions of chest compression. The median (IQR) values of SWPF at three cycles were 17,063 (10,145-26,373), 15,683 (9477-32,394), and 16,960 (7695-27,279). There was no disconnection of equipment or mechanical failures during chest compression. In addition, most participants (83%) replied that they rarely or never experienced difficulty in verbal communication and felt that the loose-fitting PAPR was comfortable.

Conclusions

The loose-fitting PAPRs provided sufficient respiratory protection without disturbances during chest compression.

Le texte complet de cet article est disponible en PDF.

Graphical abstract




Unlabelled Image

Le texte complet de cet article est disponible en PDF.

Keywords : Respiratory protective devices, Infection control, Cardiopulmonary resuscitation, Chest, Compression


Plan


© 2020  Elsevier Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 44

P. 235-240 - juin 2021 Retour au numéro
Article précédent Article précédent
  • Prehospital Shock Precautions on Triage (PSPoT) score to assess in-hospital mortality for septic shock
  • Romain Jouffroy, Jean Pierre Tourtier, Pascal Philippe, Emmanuel Bloch-Laine, Vincent Bounes, Papa Gueye-Ngalgou, Benoit Vivien
| Article suivant Article suivant
  • The representation of women and underrepresented minorities in emergency medicine: A look into resident diversity
  • Benjamin Partiali, Sandra Oska, Antonio Barbat, Adam Folbe

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.