Accuracy of ultrasound for endotracheal intubation between different transducer types - 13/12/19

Doi : 10.1016/j.ajem.2019.03.016 
Michael Gottlieb, MD a, , Dallas Holladay, DO a, Katharine Burns, MD b, Stephen R. Gore, MD a, Collin Wulff, DO a, Shital Shah, PhD a, John Bailitz, MD c
a Department of Emergency Medicine, Rush University Medical Center, Chicago, IL, United States of America 
b Department of Emergency Medicine, Advocate Christ Medical Center, Chicago, IL, United States of America 
c Department of Emergency Medicine, Feinberg School of Medicine, Northwestern Memorial Hospital, Chicago, IL, United States of America 

Corresponding author at: 1750 West Harrison Street, Suite 108 Kellogg, Chicago, IL 60612, United States of America.1750 West Harrison StreetSuite 108 KelloggChicagoIL60612United States of America

Abstract

Introduction

Ultrasound has been increasingly utilized for the identification of endotracheal tube (ETT) location after an intubation attempt, particularly among patients in cardiac arrest. However, prior studies have varied with respect to the choice of transducer and no studies have directly compared the accuracy between transducer types. Our study is the first to directly compare the accuracy of ETT confirmation between the linear and curvilinear transducer.

Methods

This study was performed in a cadaver lab using three different cadavers chosen to represent varying neck circumferences. Cadavers were randomized to tracheal or esophageal intubation. Blinded sonographers assessed the location of the ETT using either a linear or curvilinear transducer in an alternating sequence. Accuracy of sonographer identification, time to identification, and operator confidence were assessed.

Results

Four hundred and five assessments were performed with 198 (48.9%) tracheal and 207 (51.1%) esophageal intubations. The linear transducer was 98% (95% CI 95.1% to 99.2%) accurate. The curvilinear transducer was 95% (95% CI 91.1% to 97.3%) accurate. The mean time to identification was significantly lower with the linear transducer [7.46 s (95% CI 6.23 to 8.7 s)] as compared with the curvilinear transducer [11.63 s (95% CI 9.05 to 14.2 s)]. The mean operator confidence was significantly higher with the linear transducer [4.84/5.0 (95% CI 4.76 to 4.91)] than with the curvilinear transducer [4.44/5.0 (95% CI 4.3 to 4.57)]. All operators preferred the linear transducer over the curvilinear transducer.

Conclusion

The diagnostic accuracy of ultrasound for ETT confirmation did not significantly differ between ultrasound transducer types, but the curvilinear transducer was associated with a longer time to confirmation and lower operator confidence. Further studies are needed to determine if the accuracy would change with more novice providers or in specific patient populations.

Le texte complet de cet article est disponible en PDF.

Keywords : Intubation, Endotracheal, Ultrasound, Probe, Transducer, Linear, Curvilinear, Frequency


Plan


© 2019  Elsevier Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 37 - N° 12

P. 2182-2185 - décembre 2019 Retour au numéro
Article précédent Article précédent
  • New classifications for Life-threatening foreign body airway obstruction
  • Yutaka Igarashi, Tatsuya Norii, Kim Sung-Ho, Shimpei Nagata, Takashi Tagami, Jon Femling, Yasuaki Mizushima, Hiroyuki Yokota
| Article suivant Article suivant
  • Designing efficient emergency departments: Discrete event simulation of internal-waiting areas and split flow sorting
  • Benjamin Easter, Negin Houshiarian, Debajyoti Pati, Jennifer L. Wiler

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.