Noise assessment across two generations of iterative reconstruction algorithms of three manufacturers using bone reconstruction kernel - 26/11/19

Doi : 10.1016/j.diii.2019.08.003 
J. Greffier , J. Frandon, A. Larbi, D. Om, J.P. Beregi, F. Pereira
 EA 2415, Department of Radiology, Medical Imaging Group Nîmes, University of Montpellier, CHU Nîmes, boulevard du Professeur Robert Debré, 30029 Nîmes cedex, France 

Corresponding author.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Abstract

Purpose

To compare the noise-magnitude and noise-texture obtained using strong kernel across two generations of iterative reconstruction (IR) algorithms proposed by three manufacturers.

Materials and methods

Five computed tomography (CT) systems equipped with two generations of IR algorithm (hybrid/statistical IR [H/SIR] or full/partial model-based IR [MBIR]) were compared. Acquisitions on Catphan 600 phantom were performed at 120kV and three dose levels (CTDIvol: 3, 7 and 12mGy). Raw data were reconstructed using standard “bone” kernel for filtered back projection and one iterative level of two generations of IR algorithms. Contrast-to-noise ratio (CNR) was computed using three regions of interest placed semi-automatically: two placed in the low-density polyethylene and Teflon inserts and another placed on the solid water. Noise power spectrum (NPS) was computed to assess the NPS-peak and noise-texture.

Results

CNR was significantly greater in MBIR compared to H/SIR algorithms for all CT systems (P<0.01). CNR were improved on average from H/SIR to MBIR of 36±14% [SD] (range: 24–57%) for GE-Healthcare, 109±19 [SD] % (range: 89–139%) for Philips Healthcare and 42±5 [SD] % (range: 36–47%) for Siemens Healthineers. The mean NPS peak decreased from H/SIR to MBIR by −41±6 [SD] % (range: −47–−35%) for GE Healthcare, −79±3 [SD] % (range: −82–−76%) for Philips Healthcare and −52±2 [SD] % (range: −54–−51%) for Siemens Healthineers systems. NPS spatial frequencies were greater with MBIR than with H/SIR for Philips Healthcare (20 ± 2 [SD] %; range: 19–22%) and for Siemens Healthineers (9±5 [SD] %; range: 4–14%) but lower for GE Healthcare (−17±3 [SD] %; range: −14–−20%).

Conclusion

Using bone kernel with recent MBIR algorithms reduces the noise-magnitude for all CT systems assessed. Noise texture shifted towards high frequency for Siemens Healthineers and Philips Healthcare but the opposite for GE Healthcare.

Le texte complet de cet article est disponible en PDF.

Keywords : Multidetector computed tomography, Image quality enhancement, Iterative reconstruction, Optimization, Noise power spectrum


Plan


© 2019  Société française de radiologie. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 100 - N° 12

P. 763-770 - décembre 2019 Retour au numéro
Article précédent Article précédent
  • Sedation and analgesia in interventional radiology: Where do we stand, where are we heading and why does it matter?
  • F.H. Cornelis, E. Monard, M.-A. Moulin, E. Vignaud, F. Laveissiere, M. Ben Ammar, M. Nouri-Neuville, M. Barral, B. Lombart
| Article suivant Article suivant
  • Impact of an ultrasound-guided radiofrequency ablation training program on the outcomes in patients with hepatocellular carcinoma
  • R. Takai Takamatsu, A. Okano, G. Yamakawa, K. Mizukoshi, H. Obayashi, M. Ohana

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.