Assessment of focused multivector ultraviolet disinfection withshadowless delivery using 5-point multisided sampling ofpatientcare equipment without manual-chemical disinfection - 30/03/19

Doi : 10.1016/j.ajic.2018.09.019 
Donna Armellino, RN, DNP, CIC a, , Thomas J. Walsh, MD, PhD b, Vidmantas Petraitis, MD b, Wladyslaw Kowalski, PhD c
a Infection Prevention, Northwell Health, Lake Success, NY 
b Infectious Diseases Translational Research Laboratory, Weill Cornell Medicine of Cornell University, New York, NY 
c Clinical and Research Programs, PurpleSun Inc, New York, NY 

Address correspondence to Donna Armellino, RN, DNP, CIC, Infection Prevention, Northwell Health, 1985 Marcus Ave, Ste LL 106, Lake Success, NY 11042.Infection PreventionNorthwell Health,1985 Marcus Ave Ste LL 106Lake SuccessNY11042

Résumé

Background

The aim of this study was to evaluate the performance of a focused multivector ultraviolet (FMUV) system employing shadowless delivery with a 90-second disinfection cycle for patient care equipment inside and outside the operating room (OR) suite without manual-chemical disinfection.

Methods

A 5-point multisided sampling protocol was utilized to measure the microbial burden on objects inside and outside the OR environment in a 3-phase nonrandomized observational study. Surface sampling was performed pre- and postdisinfection in between cases (IBCs) to assess the performance of manual-chemical disinfection. FMUV system performance was separately assessed pre- and postdisinfection before the first case and IBCs. Additionally, visibly clean high-touch objects were sampled outside the OR, and the microbial burden reductions after FMUV disinfection were quantified without manual-chemical disinfection.

Results

Manual-chemical disinfection reduced the active microbial burden on sampled objects IBCs by 52.8%-90.9% (P < .05). FMUV reduced the active microbial burden by 92%-97.7% (P < .0001) before the firstcase and IBCs combined, and 96.3%-99.6% (P < .0001) on objects outside the OR without chemical disinfection.

Conclusions

Five-point multisided sampling proved effective for assessing disinfection performance on all exterior sides of equipment. FMUV produced significant overall reductions of the microbial burden on patient care equipment in all study phases and independent of manual cleaning and chemical disinfection.

Le texte complet de cet article est disponible en PDF.

Key Words : Environmental disinfection, Surface sampling techniques, Disinfectants, Manual cleaning, Infection prevention, Ultraviolet germicidal irradiation


Plan


 Funding/support: This study was supported by Northwell Ventures.
 PurpleSun Inc received grant support from Health 2.0 and New York City Economic Development Corporation.
 Conflicts of interest: PurpleSun is a portfolio company of Northwell Ventures.
 T.J.W. and V.P. are paid advisors to PurpleSun, and the fees will go to Weill Cornell Medicine.


© 2018  Publié par Elsevier Masson SAS.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 47 - N° 4

P. 409-414 - avril 2019 Retour au numéro
Article précédent Article précédent
  • A trial of pulsed xenon ultraviolet disinfection to reduce Clostridioides difficile infection
  • Priya Sampathkumar, Carla Folkert, Jean E. Barth, Larry Nation, Michel Benz, Allison Hesse, Catherine L. Mielke MS,, Kathryn W. Zaveleta
| Article suivant Article suivant
  • Predictors of methicillin-resistant Staphylococcus aureus and vancomycin-resistant enterococci co-colonization among nursing facility patients
  • Kevin Heinze, Mohammed Kabeto, Emily Toth Martin, Marco Cassone, Liam Hicks, Lona Mody

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2025 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.