3.0 TESLA CONTRAST-ENHANCED MR ANGIOGRAPHY OF CAROTID ARTERY STENTS: IN VITRO MEASUREMENTS AS COMPARED WITH 1.5 TESLA - 21/03/08

Doi : JNR-04-2006-33-2-0150-9861-101019-200601689 

S. Hähnel,

T.H. Nguyen-Trong,

S. Rohde,

M. Hartmann,

C. Braun,

K. Sartor,

S. Heiland

Voir les affiliations

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Étude in vitro des stents carotides en ARM injectée à 3 Tesla et à 1.5 Tesla

Objectif : Évaluer l’aspect des stents carotides en ARM injectée à 3 et 1.5 Tesla.

Méthodes : L’aspect de 19 stents (Acculink-Guidant, Dynalink-Guidant, Neuroform-Boston Scientific, Omnilink-Guidant, Protege EV3, Carotid wallstent-Boston Scirntific, Xact Abbott) de matériaux et dimensions (4,0 mm à 10,0 mm) différentes a été étudié en ARM injectée à 3.0 et 1.5 Tesla. Pour chaque stent, le rétrécissement artificiel de la lumière (RAL) a été mesuré à partir d’une comparaison de rapports contraste sur bruit du signal résiduel.

Résultats : Dans deux cas seulement (Omnilink 7,0 mm et Omnilink 10,0 mm) le RAL était plus important à 3.0 Tesla. Dans tous les autres cas, le RAL était le même. Contrairement aux stents ferromagnétiques où le RAL était supérieur à 85 %, dans la plupart des stents en nitinol, (Acculink, Dynalink, Neuroform, Protege), le RAL était inférieur à 35 %. Dans les stents Xact, le RAL était en général de 100 % à 1.5 Tesla et variait de 31.8 à 100 % à 3Tesla.

Conclusion : L’ARM injectée après stenting carotide est considérablement altérée par un RAL aussi bien à 1.5 qu’à 3 Tesla. Néanmoins, l’ARM injectée est une méthode adaptée aux stents en nitinol quelque soit le champ. Les industriels doivent être avertis des potentiels artefacts causés par leur stents lors des examens d’imagerie non invasive telle que l’ARM injectée.

Background and Purpose: To assess the appearance of carotid artery stents at 3.0Tesla contrast enhanced magnetic resonance angiography (CE-MRA) as compared with 1.5Tesla.

Methods: 19 stents (GUIDANT Acculink, GUIDANT Dynalink, BOSTON SCIENTIFIC SMART Neuroform, GUIDANT Omnilink, EV3 Protege, BOSTON SCIENTIFIC Carotid Wallstent, ABBOTT Xact) of different materials (nitinol, stainless steal, cobalt alloy) and different sizes (4.0mm-10.0mm) were investigated regarding their appearance on CE-MRA at 3.0Tesla and at 1.5 Tesla. For each stent artificial lumen narrowing (ALN) was calculated based on a pixel-by-pixel profile of the contrast-to-noise-ratio giving an objective indicator for the size of the evaluable stent diameter.

Results: Only in two stents (Omnilink 7.0mm, Omnilink 10.0mm) was ALN higher at 3.0Tesla relative to 1.5Tesla. In all other stents ALN at 3.0Tesla was the same or even lower as compared with 1.5Tesla. In contrast to the ferromagnetic stents where ALN was typically higher than 85%, in most of the nitinol stents (Acculink, Dynalink, Neuroform, Protege) ALN was below 35%. In the Xact stents ALN was generally 100% at 1.5Tesla and ranged between 31.8% and 100% at 3.0Tesla.

Conclusion: CE-MRA after carotid artery stenting is considerably impaired by ALN both at 1.5Tesla and at 3.0Tesla. Nevertheless, CE-MRA is well suited for the examination of carotid artery stents made of nitinol at both field strengths. Stent manufacturers should be aware of potential artifacts caused by their stents during noninvasive diagnostic methods such as CE-MRA.


Mots clés : angiographie par résonance magnétique , stenting , artefacts , artère carotide

Keywords: Magnetic Resonance Angiography , Stenting , Artifacts , Carotid Artery


Plan



© 2006 Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 33 - N° 2

P. 75-80 - avril 2006 Retour au numéro

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.