Diagnostic accuracy of fourth-generation ARCHITECT HIV Ag/Ab Combo assay and utility of signal-to-cutoff ratio to predict false-positive HIV tests in pregnancy - 31/07/18

Doi : 10.1016/j.ajog.2018.06.008 
Emily H. Adhikari, MD a, , Devin Macias, MD a, Donna Gaffney, MT(ASCP)SM b, Sarah White, MD a, Vanessa L. Rogers, MD a, Donald D. McIntire, PhD a, Scott W. Roberts, MD a
a Department of Obstetrics and Gynecology, University of Texas Southwestern Medical Center, Dallas, TX 
b Clinical Microbiology Laboratory, Parkland Health and Hospital System, Dallas, TX 

Corresponding author: Emily Adhikari, MD.
Sous presse. Épreuves corrigées par l'auteur. Disponible en ligne depuis le Tuesday 31 July 2018

Abstract

Background

False-positive HIV screening tests in pregnancy may lead to unnecessary interventions in labor. In 2014, the Centers for Disease Control and Prevention released a new algorithm for HIV diagnosis using a fourth-generation screening test, which detects antibodies to HIV as well as p24 antigen and has a shorter window period compared with prior generations. A reactive screen requires a differentiation assay, and supplemental qualitative RNA testing is necessary for nonreactive differentiation assay. One screening test, the ARCHITECT Ag/Ab Combo assay, is described to have 100% sensitivity and >99% specificity in nonpregnant populations; however, its clinical performance in pregnancy has not been well described.

Objective

The objective of the study was to determine the performance of the ARCHITECT assay among pregnant women at a large county hospital and to assess whether the relative signal-to-cutoff ratio can be used to differentiate between false-positive vs confirmed HIV infections in women with a nonreactive differentiation assay.

Study Design

This is a retrospective review of fourth-generation HIV testing in pregnant women at Parkland Hospital between June 1, 2015, and Jan. 31, 2017. We identified gravidas screened using the ARCHITECT Ag/Ab Combo assay (index test), with reflex to differentiation assay. Women with reactive ARCHITECT and nonreactive differentiation assay were evaluated with a qualitative RNA assay (reference standard). We calculated sensitivity, specificity, predictive value, and false-positive rate of the ARCHITECT screening assay in our population and described characteristics of women with false-positive HIV testing vs confirmed infection. Among women with a nonreactive differentiation assay, we compared interventions among women with and without a qualitative RNA assay result available at delivery and examined relative signal-to-cutoff ratios of the ARCHITECT assay in women with false-positive vs confirmed HIV infection.

Results

A total of 21,163 pregnant women were screened using the ARCHITECT assay, and 190 tested positive. Of these, 33 of 190 (17%) women had false-positive HIV screening tests (28 deliveries available for analysis), and 157 of 190 (83%) had confirmed HIV-1 infection (140 available for analysis). Diagnostic accuracy of the ARCHITECT HIV Ag/Ab Combo assay in our prenatal population (with 95% confidence interval) was as follows: sensitivity, 100% (97.7–100%); specificity, 99.8% (99.8–99.9%); positive likelihood ratio, 636 (453–895); negative likelihood ratio, 0.0 (NA); positive predictive value, 83% (77–88%); and false positive rate, 0.16% (0.11–0.22%), with a prevalence of 7 per 1000. Women with false-positive HIV testing were younger and more likely of Hispanic ethnicity. A qualitative RNA assay (reference standard) was performed prenatally in 24 (86%) and quantitative viral load in 22 (92%). Interventions occurred more frequently in women without a qualitative RNA assay result available at delivery, including intrapartum zidovudine (75% vs 4%, P = .002), breastfeeding delay (75% vs 8%, P = .001), and neonatal zidovudine initiation (75% vs 4%, P = .002). The ARCHITECT signal-to-cutoff ratio was significantly lower for women with false-positive HIV tests compared with those with established HIV infection (1.89 [1.27, 2.73] vs 533.65 [391.12, 737.22], respectively, P < .001).

Conclusion

While the performance of the fourth-generation ARCHITECT HIV Ag/Ab Combo assay among pregnant women is comparable with that reported in nonpregnant populations, clinical implications of using a screening test with a positive predictive value of 83% in pregnancy are significant. When the qualitative RNA assay result is unavailable, absence of risk factors in combination with an ARCHITECT HIV Ag/Ab assay S/Co ratio <5 and nonreactive differentiation assay provide sufficient evidence to support deferral of unnecessary intrapartum interventions while awaiting qualitative RNA results.

Le texte complet de cet article est disponible en PDF.

Key words : false positive HIV tests in pregnancy, fourth-generation HIV screening


Plan


 The authors report no conflict of interest.
 Cite this article as: Adhikari EH, Macias D, Gaffney D, et al. Diagnostic accuracy of fourth-generation ARCHITECT HIV Ag/Ab Combo assay and utility of signal-to-cutoff ratio to predict false-positive HIV tests in pregnancy. Am J Obstet Gynecol 2017;xxx:xx-xx.


© 2018  Elsevier Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.