Emergency medical technician-performed point-of-care blood analysis using the capillary blood obtained from skin puncture - 22/06/18

Doi : 10.1016/j.ajem.2017.12.025 
Changsun Kim, MD, PhD , 1 , Hansol Kim, MD 1
 Department of Emergency Medicine, Hanyang University College of Medicine, Seoul, Republic of Korea 

Corresponding author at: Department of Emergency Medicine, Hanyang University Guri Hospital, 153, Gyeongchun-ro(st), Guri-si, GyeongGi-do 471-701, Republic of Korea.Department of Emergency MedicineHanyang University Guri Hospital153, Gyeongchun-ro(st)Guri-siGyeongGi-do471-701Republic of Korea

Abstract

Objective

Comparing a point-of-care (POC) test using the capillary blood obtained from skin puncture with conventional laboratory tests.

Methods

In this study, which was conducted at the emergency department of a tertiary care hospital in April–July 2017, 232 patients were enrolled, and three types of blood samples (capillary blood from skin puncture, arterial and venous blood from blood vessel puncture) were simultaneously collected. Each blood sample was analyzed using a POC analyzer (epoc® system, USA), an arterial blood gas analyzer (pHOx®Ultra, Nova biomedical, USA) and venous blood analyzers (AU5800, DxH2401, Beckman Coulter, USA). Twelve parameters were compared between the epoc and reference analyzers, with an equivalence test, Bland-Altman plot analysis and linear regression employed to show the agreement or correlation between the two methods.

Results

The pH, HCO3, Ca2+, Na+, K+, Cl, glucose, Hb and Hct measured by the epoc were equivalent to the reference values (95% confidence interval of mean difference within the range of the agreement target) with clinically inconsequential mean differences and narrow limits of agreement. All of them, except pH, had clinically acceptable agreements between the two methods (results within target value ≥80%). Of the remaining three parameters (pCO2, pO2 and lactate), the epoc pCO2 and lactate values were highly correlated with the reference device values, whereas pO2 was not. (pCO2: R2=0.824, y=1.411+0.877·x; lactate: R2=0.902, y=0.544+0.966·x; pO2: R2=0.037, y=61.6+0.431·x).

Conclusion

Most parameters, except only pO2, measured by the epoc were equivalent to or correlated with those from the reference method.

Le texte complet de cet article est disponible en PDF.

Keywords : Point-of-care systems, Capillary blood, Emergency medical services


Plan


© 2017  Elsevier Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 36 - N° 7

P. 1215-1221 - juillet 2018 Retour au numéro
Article précédent Article précédent
  • Emergency department length of stay for ethanol intoxication encounters
  • Lauren R. Klein, Brian E. Driver, James R. Miner, Marc L. Martel, Jon B. Cole
| Article suivant Article suivant
  • An alternative tool for triaging patients with possible acute coronary symptoms before admission to a chest pain unit
  • Kuang-Chau Tsai, Ray F. Lin, Chieh Lee, Ai-Hsien Li

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.