S'abonner

A randomized cross-over study comparing surgical cricothyrotomy techniques by combat medics using a synthetic cadaver model - 19/04/18

Doi : 10.1016/j.ajem.2017.11.062 
Steven G. Schauer, DO, MS a, b, c, , Jessie Renee D. Fernandez a, b, Jamie L. Roper, DO c, Derek Brown, MD c, Kristine L. Jeffers, MD c, Jonathan Srichandra, MD c, Neil B. Davids, MD d, Michael D. April, MD, DPhil, MSc c
a US Army Institute of Surgical Research, JBSA Fort Sam Houston, TX, United States 
b 59th Medical Wing, JBSA Lackland, TX, United States 
c Brooke Army Medical Center, JBSA Fort Sam Houston, TX, United States 
d Army Medical Department Center and School, JBSA Fort Sam Houston, TX, United States 

Corresponding author at: 3698 Chambers Pass, JBSA Fort Sam Houston, TX 78234, United States.3698 Chambers PassJBSA Fort Sam HoustonTX78234United States

Abstract

Objective

Cricothyrotomy is a complex procedure with a high rate of complications including failure to cannulate and injury to adjacent anatomy. The Control-Cric™ System and QuickTrach II™ represent two novel devices designed to optimize success and minimize complications with this procedure. This study compares these two devices against a standard open surgical technique.

Methods

We conducted a randomized crossover study of United States Army combat medics using a synthetic cadaver model. Participants performed a surgical cricothyrotomy using the standard open surgical technique, Control-Cric™ System, and QuickTrach II™ device in a random order. The primary outcome was time to successful cannulation. The secondary outcome was first-attempt success. We also surveyed participants after performing the procedures as to their preferences.

Results

Of 70 enrolled subjects, 65 completed all study procedures. Of those that successfully cannulated, the mean times to cannulation were comparable for all three methods: standard 51.0s (95% CI 45.2–56.8), QuickTrach II™ 39.8s (95% CI 31.4–48.2) and the Cric-Control™ 53.6 (95% CI 45.7–61.4). Cannulation failure rates were not significantly different: standard 6.2%, QuickTrach II™ 13.9%, Cric-Control™ 18.5% (p=0.106). First pass success rates were also similar (93.4%, 91.1%, 88.7%, respectively, p=0.670). Of respondents completing the post-study survey, a majority (52.3%) preferred the QuickTrach II™ device.

Conclusions

We identified no significant differences between the three cricothyrotomy techniques with regards to time to successful cannulation or first-pass success.

Le texte complet de cet article est disponible en PDF.

Keywords : Surgical, Airway, Military, Prehospital, Cricothyrotomy


Plan


© 2017  Publié par Elsevier Masson SAS.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 36 - N° 4

P. 651-656 - avril 2018 Retour au numéro
Article précédent Article précédent
  • Effect of platelet–lymphocyte ratio and lactate levels obtained on mortality with sepsis and septic shock
  • Ebru Biyikli, Afsin Emre Kayipmaz, Cemil Kavalci
| Article suivant Article suivant
  • Association of prehospital intubation with decreased survival among pediatric trauma patients in Iraq and Afghanistan
  • Steven G. Schauer, Jason F. Naylor, Guyon J. Hill, Allyson A. Arana, Jamie L. Roper, Michael D. April

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.