Surgical Aortic Valve Replacement Following Early Sapien-XT Valve Failure: A First - 01/03/18
Résumé |
Background |
Early degeneration of prosthetic aortic valve in transcatheter aortic valve replacement (TAVR) is a rare complication.
Method |
We report the case of a 75-year-old woman who presented with severe calcific stenosis of Edwards SAPIEN-XT valve implanted only 4 years previously. She is a Jehovah’s Witness and has background of Sjogren’s syndrome with secondary cryoglobulinaemic vasculitis. She was not a suitable candidate for valve-in-valve TAVR in view of early prosthetic valve failure by calcification.
Result |
She underwent surgical explantation of the Sapiens XT valve and mechanical aortic valve replacement.
Conclusion |
To our knowledge, this is the first time that early TAVR valve degeneration and failure of an Edwards SAPIEN-XT valve was reported which required surgical replacement with a mechanical valve.
Le texte complet de cet article est disponible en PDF.Keywords : Aortic stenosis, TAVR, Edwards Sapien-XT, Early TAVR valve degeneration
Plan
Vol 27 - N° 4
P. 513-516 - avril 2018 Retour au numéroBienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.
Déjà abonné à cette revue ?