S'abonner

Interdisciplinary Development of an Improved Emergency Department Procedural Work Surface Through Iterative Design and Use Testing in Simulated and Clinical Environments - 19/04/17

Doi : 10.1016/j.annemergmed.2016.08.436 
Xiao C. Zhang, MD, MS a, Ana M. Bermudez, BA c, Pranav M. Reddy, BS, BA c, Ravi R. Sarpatwari, MD a, Darin B. Chheng, BS d, Taylor J. Mezoian, BS d, Victoria R. Schwartz, BS d, Quinneil J. Simmons, BA d, Gregory D. Jay, MD, PhD b, e, Leo Kobayashi, MD b, f,
a Emergency Medicine Residency, Providence, RI 
b Department of Emergency Medicine, Providence, RI 
c Alpert Medical School of Brown University, Providence, RI 
d Emergency Department, Rhode Island Hospital, Providence, RI 
e School of Engineering of Brown University, Providence, RI 
f Lifespan Medical Simulation Center, Providence, RI 

Corresponding Author.

Abstract

Study objective

A stable and readily accessible work surface for bedside medical procedures represents a valuable tool for acute care providers. In emergency department (ED) settings, the design and implementation of traditional Mayo stands and related surface devices often limit their availability, portability, and usability, which can lead to suboptimal clinical practice conditions that may affect the safe and effective performance of medical procedures and delivery of patient care. We designed and built a novel, open-source, portable, bedside procedural surface through an iterative development process with use testing in simulated and live clinical environments.

Methods

The procedural surface development project was conducted between October 2014 and June 2016 at an academic referral hospital and its affiliated simulation facility. An interdisciplinary team of emergency physicians, mechanical engineers, medical students, and design students sought to construct a prototype bedside procedural surface out of off-the-shelf hardware during a collaborative university course on health care design. After determination of end-user needs and core design requirements, multiple prototypes were fabricated and iteratively modified, with early variants featuring undermattress stabilizing supports or ratcheting clamp mechanisms. Versions 1 through 4 underwent 2 hands-on usability-testing simulation sessions; version 5 was presented at a design critique held jointly by a panel of clinical and industrial design faculty for expert feedback. Responding to select feedback elements over several surface versions, investigators arrived at a near-final prototype design for fabrication and use testing in a live clinical setting. This experimental procedural surface (version 8) was constructed and then deployed for controlled usability testing against the standard Mayo stands in use at the study site ED. Clinical providers working in the ED who opted to participate in the study were provided with the prototype surface and just-in-time training on its use when performing bedside procedures. Subjects completed the validated 10-point System Usability Scale postshift for the surface that they had used. The study protocol was approved by the institutional review board.

Results

Multiple prototypes and recursive design revisions resulted in a fully functional, portable, and durable bedside procedural surface that featured a stainless steel tray and intuitive hook-and-lock mechanisms for attachment to ED stretcher bed rails. Forty-two control and 40 experimental group subjects participated and completed questionnaires. The median System Usability Scale score (out of 100; higher scores associated with better usability) was 72.5 (interquartile range [IQR] 51.3 to 86.3) for the Mayo stand; the experimental surface was scored at 93.8 (IQR 84.4 to 97.5 for a difference in medians of 17.5 (95% confidence interval 10 to 27.5). Subjects reported several usability challenges with the Mayo stand; the experimental surface was reviewed as easy to use, simple, and functional. In accordance with experimental live environment deployment, questionnaire responses, and end-user suggestions, the project team finalized the design specification for the experimental procedural surface for open dissemination.

Conclusion

An iterative, interdisciplinary approach was used to generate, evaluate, revise, and finalize the design specification for a new procedural surface that met all core end-user requirements. The final surface design was evaluated favorably on a validated usability tool against Mayo stands when use tested in simulated and live clinical settings.

Le texte complet de cet article est disponible en PDF.

Plan


 Please see page 276 for the Editor’s Capsule Summary of this article.
 Supervising editor: Daniel A. Handel, MD, MBA
 Funding and support: By Annals policy, all authors are required to disclose any and all commercial, financial, and other relationships in any way related to the subject of this article as per ICMJE conflict of interest guidelines (see www.icmje.org/). The authors have stated that no such relationships exist. This article is based on work supported by the Department of Emergency Medicine at Alpert Medical School of Brown University, Rhode Island Hospital (RIH), the Rhode Island School of Design (RISD), and the University Emergency Medicine Foundation (UEMF). The study was additionally funded by the Brown Emergency Medicine Resident Scholarly Development Award (2014 to 2015) and Resident Research Seed Grant (2015 to 2016).
 Any opinions, findings, and conclusions or recommendations expressed in this article are those of the authors and do not necessarily reflect the views of the department, RIH, RISD, or UEMF.
 A podcast for this article is available at www.annemergmed.com.


© 2016  American College of Emergency Physicians. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 69 - N° 3

P. 275-283 - mars 2017 Retour au numéro
Article précédent Article précédent
  • Information for Readers
| Article suivant Article suivant
  • Use or Abuse? A Qualitative Study of Emergency Physicians' Views on Use of Observation Stays at Three Hospitals in the United States and England
  • Graham P. Martin, Brad Wright, Azeemuddin Ahmed, Jay Banerjee, Suzanne Mason, Damian Roland

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.