S'abonner

Impact of Different Iterations of Devices and Degree of Aortic Valve Calcium on Paravalvular Regurgitation After Transcatheter Aortic Valve Implantation - 04/08/16

Doi : 10.1016/j.amjcard.2016.05.053 
William K.F. Kong, MD a, b, Philippe J. van Rosendael, MD a, Frank van der Kley, MD a, Arend de Weger, MD c, Vasileios Kamperidis, MD, PhD a, Madelien V. Regeer, MD a, Nina Ajmone Marsan, MD, PhD a, Jeroen J. Bax, MD, PhD a, Victoria Delgado, MD, PhD a,
a Department of Cardiology, Leiden University Medical Center, Leiden, the Netherlands 
c Department of Cardiothoracic surgery, Leiden University Medical Center, Leiden, the Netherlands 
b Department of Cardiology, National University Heart Centre, Singapore, Singapore 

Corresponding author: Tel: (+31) 71-526-2020; fax: (+31) 71-526-6809.

Abstract

The iterations of the SAPIEN prosthesis might impact the incidence and grade of paravalvular regurgitation (PVR). The aim of this study was to assess the impact of iterations of balloon-expandable valves (SAPIEN, SAPIEN XT, and SAPIEN 3) and degree of aortic valve calcification (AVC) on the severity of PVR after transcatheter aortic valve implantation (TAVI). Comprehensive echocardiographic examinations and multidetector computed tomography (MDCT) were performed in 272 patients (127 men, 81 ± 7 years old, logistic EuroScore of 21 ± 13%) who underwent TAVI with 23- and 26-mm balloon-expandable valves. The degree of AVC was assessed with MDCT. PVR grade was assessed with echocardiography. The cover index was calculated as (prosthesis area − MDCT annulus area)/prosthesis area. SAPIEN, SAPIEN XT, and SAPIEN 3 prostheses were implanted in 103 patients (38%), 105 patients (38.5%), and 64 patients (23.5%), respectively. Significant PVR (≥moderate) occurred in 14%, 10%, and 0% of patients receiving the SAPIEN, SAPIEN XT, and SAPIEN 3, respectively (p = 0.010). Across the groups, the aortic annulus size, degree of calcification, and cover index were comparable. Larger burden of AVC was independently associated with significant PVR (odds ratio 3.48, p = 0.006) after adjusting for age, body surface area, gender, aortic annulus area, cover index, and prosthesis iteration. SAPIEN 3 was associated with lower frequency of significant PVR (odds ratio 0.31, p = 0.002). In conclusion, the incidence of significant PVR significantly decreased over time with improvement in valve design. SAPIEN 3 was associated with less significant PVR after TAVI independently of the AVC burden.

Le texte complet de cet article est disponible en PDF.

Plan


 The Department of Cardiology received grants from Edwards Lifesciences, Biotronik, Medtronic, Boston Scientific Corporation.
 See page 571 for disclosure information.


© 2016  Elsevier Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 118 - N° 4

P. 567-571 - août 2016 Retour au numéro
Article précédent Article précédent
  • Transcatheter Aortic Valve Replacement Versus Surgery in Women at High Risk for Surgical Aortic Valve Replacement (from the CoreValve US High Risk Pivotal Trial)
  • Kimberly A. Skelding, Steven J. Yakubov, Neal S. Kleiman, Michael J. Reardon, David H. Adams, Jian Huang, John K. Forrest, Jeffrey J. Popma
| Article suivant Article suivant
  • Comparison of Results of Transcatheter Aortic Valve Implantation in Patients With Versus Without Active Cancer
  • Yusuke Watanabe, Ken Kozuma, Hirofumi Hioki, Hideyuki Kawashima, Yugo Nara, Akihisa Kataoka, Shinichi Shirai, Norio Tada, Motoharu Araki, Kensuke Takagi, Futoshi Yamanaka, Masanori Yamamoto, Kentaro Hayashida

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.