Abstract

Background

Blood metal ion measurements have become a cornerstone in the surveillance of metal-on-metal (MoM) hip replacements. Interpretation of these levels in patients with bilateral MoM hip replacements, however, is challenging. We therefore asked (1) if there is a clinically significant change in whole blood (WB) Co or Cr levels in repeated WB assessment in patients operated on with bilateral ASR hip replacements, and (2) what proportion of patients has WB Co or Cr level below the previously established safe upper limits (SUL) (Co<5.0μg/L, Cr<7.4μg/L) in the repeated WB metal ion assessment.

Hypothesis

We hypothesized that there is a significant difference between repeated Co and Cr level measurement performed within one year±4 months interval in patients with bilateral ASR hip replacements.

Materials and methods

We identified all patients (n=139) who had received bilateral ASR hip replacements (278 hips). Patients (n=76, 152 hips) who had undergone two blood metal ion measurements within eight to sixteen months’ time interval were included in the study analysis. Study cohort included 38 patients with bilateral ASR hip resurfacings (mean age 53 years, median follow-up 4.6 years, median femoral diameter 51mm) and 38 patients with bilateral ASR XL total hip replacements (mean age 60 years, median follow-up 3.6 years, median femoral diameter 49mm).

Results

There was no significant change in Co and Cr levels between the repeated measurements in the HR cohort. Both Co and Cr were significantly higher in the THR cohort in the repeated measurement (Co: 8.3μg/L vs. 12.6μg/L, Cr: 3.15μg/L vs. 3.4μg/L, both P<0.001). WB Co and Cr levels remained below SUL and within their initial values during a mean one-year measurement interval in the majority of patients with bilateral high risk HR device. In contrast to this, 21.1% of patients with THRs had WB Co ion levels exceeding the SUL in the first measurement. WB Co values significantly increased over the measurement interval in the THR group.

Discussion

Annual blood metal ion measurement is not useful in patients with bilateral hip resurfacings who have undergone at least one preliminary screening including both metal ion measurement and cross-sectional imaging. It seems, however, to be beneficial in patients with bilateral high risk MoM THR.

Level of evidence

Level III (retrospective comparative study).

Le texte complet de cet article est disponible en PDF.

Keywords : Metal-on-metal, Hip resurfacing, Metal ion levels, Surveillance


Plan


© 2016  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 102 - N° 2

P. 167-173 - avril 2016 Retour au numéro
Article précédent Article précédent
  • Can the presence of an infection be predicted before a revision total hip arthroplasty? Preliminary study to establish an infection score
  • J.-Y. Jenny, B. Adamczewski, E. De Thomasson, J. Godet, H. Bonfait, C. Delaunay, French Hip, Knee Society (SFHG)
| Article suivant Article suivant
  • Effects of third fragment size and displacement on non-union of femoral shaft fractures after locking for intramedullary nailing
  • J.R. Lee, H.-J. Kim, K.-B. Lee

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.