S'abonner

A new chest compression depth indicator would increase compression depth without increasing overcompression risk - 05/12/15

Doi : 10.1016/j.ajem.2015.08.014 
Che-uk Lee, MD 1, Ji Eun Hwang, MD 1, Joonghee Kim, MD, MS , 1 , Joong Eui Rhee, MD, PhD 2, Kyuseok Kim, MD, PhD 1, Taeyun Kim, MD, PhD 1, You Hwan Jo, MD, PhD 1, Jae Hyuk Lee, MD, PhD 1, Yu Jin Kim, MD, PhD 1, Jae Yun Jung, MD, PhD 1
 Department of Emergency Medicine, Seoul National University Bundang Hospital, Seongnam-si, Gyeonggi-do, 463-707, Republic of Korea 

Corresponding author at: Department of Emergency Medicine, Seoul National University, Bundang Hospital, 166 Gumi-ro, Bundang-gu, Seongnam-si, Gyeonggi-do, 463-707, Republic of Korea. Tel.: +82 31 787 7578; fax: +82 31 787 4055.Department of Emergency Medicine, Seoul National University, Bundang Hospital166 Gumi-ro, Bundang-guSeongnam-siGyeonggi-do463-707Republic of Korea

Abstract

Purpose

Adequate chest compression (CC) depth is critical for effective cardiopulmonary resuscitation. Pediatric resuscitation guidelines recommend that CC be at least one-third of the anterior-posterior (AP) chest diameter or approximately 4 cm in infants and 5 cm in children. We aimed to find a better indicator of CC depth that maximizes CC depth while also minimizing injury.

Basic procedures

Chest computed tomographic images of patients aged 8 years and younger were measured for external diameter (ED) (AP distance from skin to skin) and internal diameter (AP distance between internal surface of anterior chest wall and anterior surface of vertebral body) at the midway of the lower half of the sternum. Compressible depth was defined as 1 cm short of internal diameter. We determined that up to a 10% estimated risk of overcompression is acceptable and approximated a quantile regression line for the 10th percentile of compressible depth on ED. After rounding coefficients, we used its equation as a new indicator.

Main findings

A total of 426 images were analyzed. The new indicator had a slope of 0.5 and an intercept of −1.9 cm (1 fingerbreadth). Compared to one-third ED, the new indicator would provide deeper CC with average difference of 1.9 mm (95% confidence interval, 1.6-2.2 mm) without increasing the risk of overcompression (both 4.9%). Chest compression of 4/5 cm would provide deeper CC compared to the new indicator (difference, 3.5 mm; 95% confidence interval, 2.7-4.1 mm); however, its overcompression risk was too high (31.5%).

Principal conclusion

Chest compression of one-half ED minus 1 fingerbreadth maximizes CC depth without increasing overcompression in pediatric population.

Le texte complet de cet article est disponible en PDF.

Plan


 Prior presentations: None.
☆☆ Funding sources: None.
 Conflicts of interest: All authors do not have any conflicts of interest.


© 2015  Elsevier Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 33 - N° 12

P. 1755-1759 - décembre 2015 Retour au numéro
Article précédent Article précédent
  • Femur fractures should not be considered distracting injuries for cervical spine assessment
  • Robert T. Dahlquist, Peter E. Fischer, Harsh Desai, Amelia Rogers, A. Britton Christmas, Michael A. Gibbs, Ronald F. Sing
| Article suivant Article suivant
  • Association of educational level with delay of prehospital care before reperfusion in STEMI
  • Ju Yeon Heo, Ki Jeong Hong, Sang Do Shin, Kyoung Jun Song, Young Sun Ro

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.